Opis
Neurodyn 10 Channels Description
NEURODYN 10 CHANNELS V3.0 uses microcomputer technology, i.e, it is microcontrolled. It was designed according to existing technical standards for the development of medical equipment (NBR IEC 60601-1, IEC 60601-1-2 NBR and NBR IEC 60601-2-10). NEURODYN 10 CHANNELS V3.0 is a complete unit of simple operation, with 10 output channels, allowing multiple treatment options, making it essential for clinical practice. It is an equipment for the production of average frequency current, Russian current and Aussie current for neurological disorders treatment , orthopedics, traumatology and esthetic (Eletrolipólise, Lymphatic Drainage, etc.).
Essential Performance : NEURODYN 10 CHANNELS V3.0 is a device for applying electrical current through electrodes in direct contact with the patient. This is a transcutaneous neuromuscular stimulator. NEURODYN 10 CHANNELS V3.0 was developed taking into account the need of workers who work in different areas of physical rehabilitation, medical esthetics and esthetics. It is an electro stimulator with 10 output channels for the patient, with independent control of intensity, that is, each channel has an individual control of intensity, which can be used both in a continuous, synchronous, reciprocal, unblocking, sequentially way. NEURODYN 10 CHANNELS V3.0 generates Russian current at a frequency of 2.500 Hz, Aussie current (also called the Australian current ) at a frequency of 4.000 Hz or 1.000 Hz modulated bursts with a duration of 4 ms or 2 ms, lymphatic drainage with the possibility of choosing from 4 to10 channels sequentially and eletrolipolise to treat localized fat. The application of stimulation current is via transcutaneous electrodes (Russian and Aussie current ) or needles (eletrolipolise).
The therapeutical technique of Russian or Aussie current is the application of mild electrical stimulation through electrodes placed on the body parts. The Eletrolipolise technique involves the application of gentle stimulation through needle type electrodes. The technique with the Russian or Aussie current is noninvasive and Eletrolipolise technique is invasive. Both do not have systemic effects, they are not addictive and have no side effects. The intensity current required for the treatment depends on the patient’s sensitivity. Therefore, the treatment should be started with minimum levels of intensity (very low), increasing thoroughly until they until they get the proper effect to the procedure and according to the patient’s report. When a person is subjected to various types of chain, it will feel a tingling sensation at the site or in areas between the electrodes.
This feeling is generally agreable for most of the people. The sensation is controlled by adjusting the parameters (control) of the equipment. Because of the technology used is the same as microcomputers, these controls operate through a touching keyboard. All information related to the parameters chosen by the therapist will be displayed in an alphanumeric LCD .
NEURODYN 10 CHANNELS V3.0 provides the following stimulating modes:
RUSSIAN Current (2.500Hz):
– CONTINUOUS MODE;
– SYNCHRONIZED MODE (with ramps: On, Off, Rise and Decay);
– RECIPROCAL MODE (with ramps: On, Off, Rise and Decay);
– SEQUENTIAL MODE (with option to select channels 4-10);
– CLEARING WAY.
AUSSIE Current ( AUSTRALIAN Current) (frequency of 1000 Hz or 4000 Hz with burst
length of 4 ms or 2 ms):
– CONTINUOUS MODE;
– SYNCHRONIZED MODE (with ramps: On, Off, Rise and Decay);
– RECIPROCAL MODE (with ramps: On, Off, Rise and Decay);
– SEQUENTIAL MODE (with the option to select channels 4-10);
– CLEARING WAY.
ELETROLIPOLISE (2,500 Hz modulated by 75 Hz):
– CONTINUOUS MODE:
The device allows the choice of the following parameters:
CHANNEL 1 – rgulates the current strength of channel 1;
CHANNEL 2 – regulates the current strength of channel 2;
CHANNEL 3 – regulates the current strength of channel 3;
CHANNEL 4 – regulates the current strength of channel 4;
CHANNEL 5 – regulates the current strength of channel 5;
CHANNEL 6 – regulates the current strength of channel 6;
CHANNEL 7 – regulates the current strength of channel 7;
CHANNEL 8 – regulates the current strength of channel 8;
CHANNEL 9 – regulates the current strength of channel 9;
CHANNEL 10 – regulates the current strength of channel 10.
TIMER – allows you to select application time from 1 to 60 minutes. At the end of time selected, an alarm will sound and it will stop to pass the current to the patient. The selected value will decrease as this time is running out.
Type of current:
RUSSIAN Current – Mode: CONT. (Continued):
FREQ. (Hz): – in this case the equipment is operating at an average frequency of 2,500 Hz modulated by the low frequency band (chosen by the therapist) from 1 Hz to 120 Hz The sensation of stimulation is continuous, constant.
RUSSIAN Current – Mode: SYNC or REC. (Synchronous or reciprocal):
FREQ. (Hz): – In this case the equipment is operating at an average frequency of 2,500 Hz modulated by the low frequency band (chosen by the therapist) from 1 Hz to 120 Hz the sensation of stimulation follows the chosen time on the slopes On, Off, Rise and Decay.
RISE (rising ramp pulse): – Pulse rise time, variable from 1 to 20 seconds.It regulates the speed of contraction, i.e the time from the beginning until the maximum muscle contraction. High time produces a slow but gradual contraction. Short time produces a sudden relaxation.
DECAY (ramp-down pulse) – lowering the pulse, ranging from 1 to 20 seconds.It regulates the speed at which the contraction decreases, ie the time from maximum contraction to muscle relaxation. High time produces a slow relaxation. Low time produces a sudden relaxation.
ON TIME (time on): – time of maximum muscle contraction, variable 10-60 seconds. It sets the time that current flows through the electrodes during each cycle of stimulation.
OFF TIME (time off): – resting time of muscle contraction, variable 10-60 seconds. It sets the time that no current flows through the electrodes during each cycle.
Note: When selected Russia CONTINUOUS, the parameters Rise, On, Off and Decay will be disabled. The sensation will be continuous, constant. When selected, RUSSIAN SYNCHRONIZED channels 1-10 work together at the same time, i.e, all channels running simultaneously the time chosen for Rise, On, Off and
Decay. When selected RUSSIAN RECIPROCAL, channels 1, 2, 3, 4 and 5 alternately work with
the channels 6, 7, 8, 9 and 10.
RUSSIAN Current – Mode: SEQ. (Sequential) FREQ. (Hz): – In this case the equipment is operating at an average frequency of 2,500 Hz modulated by the low frequency band (chosen by the therapist) from 1 Hz to 120 Hz.The stimulation sensation follows the chosen time on the ramp Rise.
RISE (rising ramp pulse): – Pulse rise time, variable 10-20 seconds. It regulates the speed of the channel sequence.
Note: In SEQ mode. The sensation is made from one channel to another sequentially, with the first channel only stops the passage of the third stimulus when it is passing. So there is no possibility of backflow of liquid. The device permits to choose from 4 to10 channels.
RUSSIAN Current – Mode: DES. (Clearance):
FREQ. (Hz): – In this case, the equipment is operating at an average frequency of 2,500 Hz modulated by the low frequency band (chosen by the therapist) from 1 Hz to 120 Hz. The stimulation sensation follows the chosen time on the ramp Rise.
RISE (rising ramp pulse): – Pulse rise time, variable from 1 to 20 seconds. It regulates the speed of the channel sequence .
Note: In DES mode. the sensation is made from one channel to another sequentially increasing, and soon followed by sequentially decreasing. The device offers a choice of clearing from 4 to10 channels.
AUSSIE Current ( Australian current) – Mode: CONT. (Continued):
FREQ. (Hz): – In this case the equipment is operating at an average frequency of 1000 Hz or 4000 Hz (chosen by the therapist) modulated by the low frequency band (chosen by the therapist) from 1 Hz to 120 Hz. It is possible the selection of trains pulse (burst) in 4 ms or 2 ms. The sensation of stimulation is continuous, constant.
AUSSIE Current (Australian current ) – Mode: SYNC or REC. (Synchronous or reciprocal):
FREQ. (Hz): – In this case the equipment is operating at an average frequency of 1000 Hz or 4000 Hz (chosen by the therapist) modulated by the low frequency band (chosen by the therapist) from 1 Hz to 120 Hz. It is possible the selection of pulse trains (burst) in 4 ms or 2 ms.The sensation of stimulation follows the chosen time on the slopes On, Off, Rise and Decay.
RISE (rising ramp pulse): – Pulse rise time, variable from 1to 20 seconds. It regulates the speed of contraction, i.e the time from the beginning until the maximum muscle contraction. High time produces a slow but gradual contraction.). Low time produces a sudden relaxation DECAY (ramp-down pulse) – Pulse lowering time, variable from 1 to 20 seconds. It regulates the speed at which the contraction decreases, i.e the time from maximum contraction to muscle relaxation. High time produces a slow relaxation. Low time produces a sudden relaxation.
ON TIME (time on): – Time of maximum muscle contraction, variable from 1 to 60 seconds. It sets the time that current flows through the electrodes during each cycle of stimulation.
OFF TIME (time off): – Resting time of muscle contraction, ranging from 1 to 60 seconds. It sets the time that no current flows through the electrodes during each cycle.
Note: When selected CONTINUOUS AUSSIECURRENT , parameters Rise, On, Off and Decay will be disabled. So the sensation is continuous, constant. When selected SYNCHRONIZED AUSSIE CURRENT, the channels 1-10 work together at the same time, i.e, all channels running simultaneously the time chosen for Rise, On, Off and Decay. When selected RECIPROCAL AUSSIE CURRENT, channels 1, 2, 3, 4 and 5 alternately work with the channels 6, 7, 8, 9 and 10.
AUSSIE current ( Australian current) – Mod: SEQ. (Sequential)
FREQ. (Hz): – In this case, the equipment is operating at an average frequency of 1000 Hz or 4000 Hz (chosen by the therapist) modulated by the low frequency band (chosen by the therapist) from 1 Hz to 120 Hz. It is possible to select pulse trains (burst) in 4 ms or 2 ms.The sensation of stimulation follows the chosen time on the ramp Rise.RISE (rising ramp pulse): – Pulse rise time, variable from 1 to 20 seconds.It
regulates the speed of the sequence of channels.
Note: In SEQ mode. the sensation is made from one channel to another sequentially, and the first channel only stops the passage of stimulus when the third is passing. So there is no possibility of backflow of liquid. The device offers a choice sequence from 4 to 10 channels.
AUSSIE current (Australian current ) – Mode: DES. (Clearance):
FREQ. (Hz): – In this case the equipment is operating at an average frequency of 1000 Hz or 4000 Hz (chosen by the therapist) modulated by the low frequency band (chosen by the therapist) from 1 Hz to 120 Hz. It is possible the selection of pulse trains (burst) in 4 ms or 2 ms.The sensation of stimulation follows the selected time on the ramp Rise.RISE (rising ramp pulse) – Pulse rise time, variable from 1 to 20 seconds. It regulates the speed of the sequence of channels.
Note: In DES mode. the sensation is made from one channel to another sequentially increasing, and soon followed by sequentially decreasing. The device offers a choice of clearing from 4 to10 channels.
ELETROLIPOLISE – Mode: Only CONT. (Continued):
FREQ. (Hz): In this case the equipment is operating at an average frequency of 2,500 Hz modulated by a low frequency of 75 Hz. The sensation of stimulation is continuous, constant.
PRODUCT FEATURES
NEURODYN 10 CHANNELS V3.0 is an equipment designed for continuous operation mode. It uses microcontrolled technology to ensure the accuracy of the values shown. The accuracy of operation is in accordance with the provisions of the standards for particular safety equipment for neuromuscular stimulation – NBR IEC 60601-2-10, Clause 50 / subclauses 50.1 and 50.2. The amplitude control continuously monitors the output current intensity from the minimum to the maximum and minimum value does not exceed 2% of the maximum position. The parameters such as waveform output, pulse duration, pulse repetition frequency, amplitude range of output current do not differ by more than 30% mentioned in the following technical description.
The values of the durations of pulses and frequency of repetition of pulses described here were measured at 50% of the maximum amplitude of the output.These parameters are for a load impedance of 1000 O. The effect of load impedance parameters described is very important. If the unit is operated outside the load impedance range specified, there may be inaccuracy in the values of the parameters, as well as modification of the waveforms described here.
NEURODYN 10 CHANNELS V3.0 is a Class II, single-phase equipment with Type BF applied part of security and protection.
Power source: – Bivolt 100/240 V ~ (50/60 Hz);
Input Power – Consumption (Max): – 50 Va;
Output channels: – 10 independent channels in amplitude: Channels from1 to10 are intended to Russian current , Aussie, and Eletrolipolise current .
Amplitude Range (mA peak to peak):
– Russian: – 0-140 mA;
– Aussie: – 0-140 mA;
– Eletrolipolise: – 0-20 mA;
Pulse Shape: – Sinusoidal (depolarized);
Pulse repetition frequency (R):
– Russian Current: – Carrying 2500 Hz modulated by low frequency choice 1-120 Hz;
– Aussie Current : – Carrying 1000 Hz or 4000 Hz modulated by low frequency (bursts
of 4 ms or 2 ms) with a choice of 1-120 Hz;
Length (w – T) of a single pulse of RUSSIAN current 50% of maximum amplitude: – 400 usec;
Length (w – T) of a single pulse of AUSSIE current 4,000 Hz at 50% of maximum amplitude: – 250 usec;
Length (w – T) of a single pulse of AUSSIE current 1,000 Hz at 50% of maximum amplitude: – 100 usec;
ON Time: – Variable 1-60 seconds;
OFF Time: – Variable 1-60 seconds;
RISE (Rising Time of Pulse Train): – Variable 10-20 sec;
DECAY (time Descent of Pulse Train): – Variable 10-20 sec;
Timer: – Variable from 1 to 60 minutes.
Dimensions (approx. in mm): – 360 x 320 x 120 (W x D x H);
Weight (approx. without accessories): – 2.6 kg;
Stacking up: – 5 boxes;
Temperature w / transport: – 50-50 ° C;
Ambient operating temperature: – 5-45 ° C.
Abbreviations of measurement units used:
mA = milliamperes;
Hz = hertz;
Va = volt amps;
mm = mm;
usec = us = microseconds;
s = seconds;
msec = milliseconds.
V ~ = AC voltage;
cm = centimeter;
W = W;
O = ohms.
Accessories included with the device:
– 10 Blue pad electrodes of silicone rubber 75 mm;
– 10 Green pad electrodes of silicone rubber 75 mm;
– 05 cables (each with four wires and the banana tips) for connecting to the patient with Russian and Aussie current- (ch. 10-10);
– 05 cables (each with 4 wires and alligator clips on the ends) for connecting to the patient with Eletrolipolise – (ch. 10-10);
– 10 elastic straps with Velcro;
– 01Pair of pens with/ facial stimulation;
– 02 conductive silicone rubber tips – 10 mm diameter (for Russian facial stimulation pens);
– 02 conductive silicone rubber tips -15mm diameter (for Russian facial stimulation pens );
– 01 detachable power cord;
– 01 tube gel;
– 01 instruction CD ;
– 01 Spare protection fuse ;
– 100 needles for acupuncture Size: Diameter = 0.25 mm and length = 50 mm / Anvisa record No. 10,390,380,001 (made by Dong Bang Acupuncture, Inc – South Korea)
Ps.: Using cables, electrodes and other accessories different from those specified above may result in increased emissions or decreased immunity of the equipment.
Note: The device and its features may change without notice